Novel atrioventricular sequential pacing approach using a transvenous atrial pacemaker and a leadless pacemaker: a case report.


ABSTRACT:

Background

The use of transvenous pacing leads is associated with the risk of developing tricuspid valve (TV) dysfunction. This develops through several mechanisms including the failure of leaflet coaptation or direct damage to the TV or to its sub-valvular apparatus and can result in significant tricuspid regurgitation (TR). Multiple approaches to pacemaker implantation after transvenous lead extraction (TLE) or surgical TV repair have been described. Placement of pacing leads across the TV is generally avoided in such circumstances.

Case summary

A 66-year-old woman presented with a year-long history of exertional dyspnoea, peripheral oedema, and postural neck pulsations. Her medical history included a dual-chamber pacemaker implantation for sinus node dysfunction 14 years