Reliability and agreement of the central and mid-peripheral corneal thickness measured by a new Scheimpflug based imaging.


ABSTRACT:

Background

To assess the intra-observer repeatability and inter-observer reproducibility of central corneal thickness (CCT) and mid-peripheral corneal thickness (MPCT) measurements using a new Scheimpflug imaging instrument (Scansys) and compare the agreement with the rotating Scheimpflug corneal tomographer (Pentacam HR).

Methods

The same well-trained operator performed the measuring using the two devices, after which Scansys measurements were repeated by another operator. Both instruments required three consecutive measurements per subject. Corneal thickness measurements were obtained by each instrument, including CCT, thinnest corneal thickness (TCT), pupil corneal thickness (PCT), and MPCT. Test-retest repeatability (TRT), within-subject coefficient of variation (CoV), a