Patient preferences for ambulatory blood pressure monitoring devices: Wrist-type or arm-type?


ABSTRACT:

Background

Ambulatory blood pressure monitoring (ABPM) is important in evaluating average 24-hour blood pressure (BP) levels, circadian rhythm, sleeping BP and BP variability but many patients are reluctant to use standard ABPM devices.

Methods

We compared two validated ABPM devices, the BPro tonometric wrist monitor and the A&D TM-2430 oscillometric upper arm monitor, for agreement of recordings and acceptability in 37 hypertensive patients (aged 55±9 years).

Results

Successful BP measurements were less frequent with the wrist-type than the arm-type device during the sleeping (66.3% vs. 92.9%, P <0.01) and awake periods (56.2% vs. 86.5%, P <0.01). Comparable paired readings showed no significant difference in systolic BP but diastolic BP (DBP) values were higher with