VHIT Testing of Vertical Semicircular Canals With Goggles Yield Different Results Depending on Which Canal Plane Being Tested.
Ontology highlight
ABSTRACT: Objective: The use of goggles to assess vertical semicircular canal function has become a standard method in vestibular testing, both in clinic and in research, but there are different methods and apparatus in use. The aim of this study was to determine what the cause of the systematic differences is between gain values in testing of the vertical semicircular canals with two different video head impulse test (vHIT) equipment in subjects with normal vestibular function. Study Design: Retrospective analysis of gain values on patients with clinically deemed normal vestibular function (absence of a corrective eye saccade), tested with either Interacoustics or Otometrics system. Prospective testing of subjects with normal vestibular function with the camera records the eye movemen
SUBMITTER: Wittmeyer Cedervall L
PROVIDER: S-EPMC8353365 | biostudies-literature | 2021
REPOSITORIES: biostudies-literature