Evaluation of automated molecular tests for the detection of SARS-CoV-2 in pooled nasopharyngeal and saliva specimens.


ABSTRACT:

Background

Pooling of samples for SARS-CoV-2 testing in low-prevalence settings has been used as an effective strategy to expand testing capacity and mitigate challenges with the shortage of supplies. We evaluated two automated molecular test systems for the detection of SARS-CoV-2 RNA in pooled specimens.

Methods

Pooled nasopharyngeal and saliva specimens were tested by Qiagen QIAstat-Dx Respiratory SARS-CoV-2 Panel (QIAstat) or Cepheid Xpert Xpress SARS-CoV-2 (Xpert), and the results were compared to that of standard RT-qPCR tests without pooling.

Results

In nasopharyngeal specimens, the sensitivity/specificity of the pool testing approach, with 5 and 10 specimens per pool, were 77%/100% (n = 105) and 74.1%/100% (n = 260) by QIAstat, and 97.1%/100% (n = 250) and 100