This report describes a manufacturer-independent evaluation of the diagnostic accuracy of the Elecsys SARS-CoV-2 antigen assay from Roche Diagnostics in a tertiary care setting.Methods
In this single-center study, we used nasopharyngeal swabs from 403 cases from the emergency department and intensive care unit of our hospital. The reference standard for detecting SARS-CoV-2 was the reverse-transcription polymerase chain reaction (RT-PCR) assay. Cycle threshold (Ct) values were recorded for positive RT-PCR assays. The index test was the Elecsys SARS-CoV-2 antigen assay. This electrochemiluminescence immunoassay produces results as cutoff index (COI) values, with values ≥1.00 being reported as positive.Results
Of the 403 cases, 47 showed positive results in