Lung diffusing capacity for nitric oxide measured by two commercial devices: a randomised crossover comparison in healthy adults.
Ontology highlight
ABSTRACT: In Europe, two commercial devices are available to measure combined single-breath diffusing capacity of the lung for nitric oxide (D LNO) and carbon monoxide (D LCO) in one manoeuvre. Reference values were derived by pooling datasets from both devices, but agreement between devices has not been established. We conducted a randomised crossover trial in 35 healthy adults (age 40.0±15.5 years, 51% female) to compare D LNO (primary end-point) between MasterScreen™ (Vyaire Medical, Mettawa, IL, USA) and HypAir (Medisoft, Dinant, Belgium) devices during a single visit under controlled conditions. Linear mixed models were used adjusting for device and period as fixed effects and random intercept for each participant. Difference in D