Inter-rater Reliability and Repeatability of Manual Anterior Segment Optical Coherence Tomography Image Grading in Keratoconus.


ABSTRACT:

Objectives

To determine the repeatability of corneal measurements from anterior segment optical coherence tomography (AS-OCT) images using ImageJ software in healthy eyes compared with eyes with keratoconus.

Methods

Anterior segment OCT images of 25 eyes from 14 healthy subjects and 25 eyes from 15 subjects with keratoconus between the ages of 20 and 80 years were evaluated. Two trained observers used ImageJ to measure the central corneal cross-sectional area and anterior and posterior corneal arc lengths. MedCalc statistical software was used to generate the intraclass correlation coefficient (ICC) and Bland-Altman plots (BAPs) for observer measurements.

Results

Observer measurements of the central corneal cross-sectional area and anterior and posterior corneal arc l