ABSTRACT: Objective
To evaluate the influence of target dose heterogeneity on normal tissue dose sparing for peripheral lung tumor stereotactic body radiation therapy (SBRT).Methods
Based on the volumetric-modulated arc therapy (VMAT) technique, three SBRT plans with homogeneous, moderate heterogeneous, and heterogeneous (HO, MHE, and HE) target doses were compared in 30 peripheral lung tumor patients. The prescription dose was 48 Gy in 4 fractions. Ten rings outside the PTV were created to limit normal tissue dosage and evaluate dose falloff.Results
When MHE and HE plans were compared to HO plans, the conformity index of the PTV was increased by approximately 0.08. The median mean lung dose (MLD), V5, V10, V20 of whole lung, D2%, D
SUBMITTER: Wei Z
PROVIDER: S-EPMC8404286 | biostudies-literature | 2021 Aug
REPOSITORIES: biostudies-literature