A rechargeable anti-thrombotic coating for blood-contacting devices.
Ontology highlight
ABSTRACT: Despite the potential of anti-thrombogenic coatings, including heparinized surfaces, to improve the performance of blood-contacting devices, the inevitable deterioration of bioactivity remains an important factor in device failure and related thrombotic complications. As a consequence, the ability to restore the bioactivity of a surface coating after implantation of a blood-contacting device provides a potentially important strategy to enhance its clinical performance. Here, we report the regeneration of a multicomponent anti-thrombogenic coating through use of an evolved sortase A to mediate reversible transpeptidation. Both recombinant thrombomodulin and a chemoenzymatically synthesized ultra-low molecular weight heparin were repeatedly and selectively immobilized or removed in a sequent
SUBMITTER: Ham HO
PROVIDER: S-EPMC8405571 | biostudies-literature | 2021 Sep
REPOSITORIES: biostudies-literature