SARS-CoV-2 (Covid-19) workplace temperature screening: Seasonal concerns for thermal detection in northern regions.


ABSTRACT: Workplace temperature screening has become standard practice during the SARS-CoV-2 pandemic. The objective was to determine the consistency of four temperature devices during exposure to simulated and actual environmental conditions reflective of a workplace. An infrared (IR) digital thermometer (accuracy(A)±0.2), IR laser thermometer (A±1), and thermal imaging camera (A±0.3) were used to measure forehead and tympanic (digital only) temperatures. The first experiment was conducted in a controlled simulated environment (-20 to 20 °C) with three participants (32-YOF, 27-YOM, 20-YOF). The second experiment used actual outdoor conditions (-0.48 to 45.6 °C) with two participants (32-YOF, 27-YOM). The tympanic measurement was the least impacted by environmental temperature (mean(±SD)): simulated (36.8(±0.18) °C) and actual (36.9(±0.16) °C). The thermal imaging camera had the lowest RMSE values (0.81-0.97 °C), with outdoor temperatures ranging from 0 to 45 °C. Environmental temperature influenced forehead temperature readings and required a resting period in a thermoneutral environment (5-9 min (-20 to -10 °C) to immediate (15-20 °C)).

SUBMITTER: Goggins KA 

PROVIDER: S-EPMC8407948 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

SARS-CoV-2 (Covid-19) workplace temperature screening: Seasonal concerns for thermal detection in northern regions.

Goggins Katie A KA   Tetzlaff Emily J EJ   Young Wesley W WW   Godwin Alison A AA  

Applied ergonomics 20210901


Workplace temperature screening has become standard practice during the SARS-CoV-2 pandemic. The objective was to determine the consistency of four temperature devices during exposure to simulated and actual environmental conditions reflective of a workplace. An infrared (IR) digital thermometer (accuracy(A)±0.2), IR laser thermometer (A±1), and thermal imaging camera (A±0.3) were used to measure forehead and tympanic (digital only) temperatures. The first experiment was conducted in a controlle  ...[more]

Similar Datasets

| S-EPMC9455844 | biostudies-literature
| S-EPMC9464234 | biostudies-literature
| S-EPMC7675320 | biostudies-literature
| S-BSST379 | biostudies-other
| S-EPMC9077488 | biostudies-literature
| S-EPMC10105261 | biostudies-literature
| S-EPMC3679019 | biostudies-literature
| S-EPMC9491009 | biostudies-literature
| S-EPMC4685772 | biostudies-literature
2021-05-28 | GSE175666 | GEO