Is there a difference in venous thrombosis rate in free flap anastomoses based on coupler diameter? A systematic review. Does Size Really Matter?


ABSTRACT:

Background

The adage is to use the largest anastomotic coupler device (coupler) size possible, since smaller an anastomosis might be more susceptible to thrombosis. It is unclear if this wisdom is supported by data. This study tests the hypothesis that there is no difference in the reported literature in thrombosis rate between different coupler sizes.

Methods

We searched PubMed, Embase, and the Cochrane Library. After screening 235 studies, we included 11 retrospective case-series. According to the criteria of Newcastle-Ottowa Scale, quality score ranged from 2 to 4 (out of 5) and funnel plots indicated publication bias. We included a total of 5930 coupled anastomoses. We calculated thrombosis rate per coupler diameter with exact confidence intervals (CIs). We regard non-ov