ABSTRACT: Purpose
To evaluate the ability of the new in vivo corneal indentation device (CID) to measure corneal biomechanical properties.Methods and results
In total, 186 eyes from 46 healthy subjects, 107 patients with primary open-angle glaucoma, and 33 patients with ocular hypertension were enrolled in a cross-sectional study. Measurements were performed using corneal visualization Scheimpflug technology (Corvis ST) and the CID. The deformation amplitude (DA), inward applanation time, inward applanation velocity (A1V), outward applanation time (A2T), outward applanation velocity (A2V), highest concavity time, DA ratio, max inverse radius (MIR), integrated radius, and stiffness parameter A1 were included as Corvis ST parameters, and stiffness and modulus were included as CID param