ABSTRACT: This work presents a comprehensive commissioning and workflow development process of a real-time, ultrasound (US) image-guided treatment planning system (TPS), a stepper and a US unit. To adequately benchmark the system, commissioning tasks were separated into (1) US imaging, (2) stepper mechanical, and (3) treatment planning aspects. Quality assurance US imaging measurements were performed following the AAPM TG-128 and GEC-ESTRO recommendations and consisted of benchmarking the spatial resolution, accuracy, and low-contrast detectability. Mechanical tests were first used to benchmark the electronic encoders within the stepper and were later expanded to evaluate the needle free length calculation accuracy. Needle reconstruction accuracy was rigorously evaluated at the treatment planning le