Unknown

Dataset Information

Avoiding transduction-induced heating in suspended microchannel resonators using piezoelectricity.


ABSTRACT: Calorimetry of single biological entities remains elusive. Suspended microchannel resonators (SMRs) offer excellent performance for real-time detection of various analytes and could hold the key to unlocking pico-calorimetry experiments. However, the typical readout techniques for SMRs are optical-based, and significant heat is dissipated in the sensor, altering the measurement and worsening the frequency noise. In this manuscript, we demonstrate for the first time full on-chip piezoelectric transduction of SMRs on which we focus a laser Doppler vibrometer to analyze its effect. We demonstrate that suddenly applying the laser to a water-filled SMR causes a resonance frequency shift, which we attribute to a local increase in temperature. When the procedure is repeated at increasing flow rat

SUBMITTER: Maillard D 

PROVIDER: S-EPMC8433141 | biostudies-literature | 2021

REPOSITORIES: biostudies-literature