Variation in the Measurement of Anti-Mullerian Hormone - What Are the Laboratory Issues?


ABSTRACT: Anti-Müllerian Hormone (AMH) is a 140 kDa homodimeric glycoprotein consisting of two identical subunits linked by disulphide bonds and is synthesised by the testes and ovaries. Its clinical applications are prediction of ovarian response and gonadotropin dose selection upon in vitro fertilization. In males, AMH is used to investigate sexual developmental disorders and gonadal function. AMH is commonly assayed by enzyme-linked immunosorbent assay or automated immunoassay formats that show variation between methods. This review applies fundamental chemical pathology concepts to explain the observed analytical variation of AMH measurement. We examine the lack of standardisation between AMH assays, the impact of antibody design on variable measurements, consider the analytical detection