Improving the robustness of MOLLI T1 maps with a dedicated motion correction algorithm.
Ontology highlight
ABSTRACT: Myocardial tissue T1 constitutes a reliable indicator of several heart diseases related to extracellular changes (e.g. edema, fibrosis) as well as fat, iron and amyloid content. Magnetic resonance (MR) T1-mapping is typically achieved by pixel-wise exponential fitting of a series of inversion or saturation recovery measurements. Good anatomical alignment between these measurements is essential for accurate T1 estimation. Motion correction is recommended to improve alignment. However, in the case of inversion recovery sequences, this correction is compromised by the intrinsic contrast variation between frames. A model-based, non-rigid motion correction method for MOLLI series was implemented and validated on a large database of cardiac clinical cases (n = 186). The method relies on a dedica
SUBMITTER: Delso G
PROVIDER: S-EPMC8448777 | biostudies-literature | 2021 Sep
REPOSITORIES: biostudies-literature