ABSTRACT: Aims
To compare the biometric data from partial coherence interferometry (PCI) and swept-source OCT (SS-OCT) in patients with age-related cataract and epiretinal membrane (ERM): ERM, ERM with foveoschisis and macular pseudohole.Methods
49 eyes of 49 subjects including 36 ERM, 9 ERM foveoschisis and 4 macular pseudohole were analysed to evaluate the axial length (AL) measurements and the presence of AL measurement errors, defined basing on the shape of the biometric output graphs and on the concordance of AL values between instruments. Eyes with ERM were divided in four stages according to OCT features (i.e. presence/absence of the foveal pit, presence of ectopic inner foveal layers, disrupted retinal layers).Results
The devices provided similar mean AL measurements