Unknown

Dataset Information

Evaluation of interference from hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by fifteen methods.


ABSTRACT:

Background

Hemoglobin C, D Punjab, E or S trait can interfere with hemoglobin A1c (HbA1c) results. We assessed whether they affect results obtained with 15 current assay methods.

Methods

Hemoglobin AA (HbAA), HbAC, HbAD Punjab, HbAE and HbAS samples were analyzed on 2 enzymatic, 4 ion-exchange HPLC and 9 immunoassay methods. Trinity Premier Hb9210 boronate affinity HPLC was the comparative method. An overall test of coincidence of least-squared linear regression lines was performed to determine if HbA1c results were statistically significantly different from those of HbAA samples. Clinically significant interference was defined as >6% difference from HbAA at 6 or 9% HbA1c compared to Premier Hb9210 using Deming regression.

Results

All methods showed statistically sign

SUBMITTER: Rohlfing C 

PROVIDER: S-EPMC8464520 | biostudies-literature | 2021 Nov

REPOSITORIES: biostudies-literature