The Impact of Phototherapy on the Accuracy of Transcutaneous Bilirubin Measurements in Neonates: Optimal Measurement Site and Timing.


ABSTRACT: Transcutaneous bilirubinometer devices are widely applied to assess neonatal hyperbilirubinemia. However, the optimal skin site and timing of transcutaneous bilirubin (TCB) measurements for the strongest correlation with total serum bilirubin (TSB) levels after phototherapy are still unclear. We conducted a retrospective observational study evaluating the correlation of TCB and TSB levels in neonates postphototherapy. The TCB measurements on the forehead and mid-sternum at 0 and 30 min postphototherapy were assessed by using a JM-103 bilirubinometer. Paired TCB and TSB measurements were assessed by Pearson correlation and Bland-Altman plots. We analyzed 40 neonates with 96 TSB and 384 TCB measurements. The TSB level correlated moderately with the forehead TCB level at 30 min postphotothera