Comparative Evaluation of Bruker Biotyper and ASTA MicroIDSys for Species Identification in a Clinical Microbiology Laboratory.


ABSTRACT: Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been widely used for microbial identification, because of its speed and accuracy, since its introduction to clinical microbiology laboratories. In this study, we evaluated the performance of ASTA MicroIDSys, a newly developed MALDI-TOF, and compared it with the widely used Bruker Biotyper. Microbial identification with the Bruker Biotyper system was performed by using a direct smear method and the Bruker Biotyper database (reference library version 6.0.0.0). The isolates were also tested in parallel, using the ASTA MicroIDSys system with a direct smear method and the MicroIDSys database, CoreDB v1.26. A total of 914 clinical isolates were recovered from the clinical specimens. Identical results