A bioinspired, passive microfluidic lobe filtration system.
Ontology highlight
ABSTRACT: Size-based microfluidic filtration systems can be affected by clogging, which prevents their use in high-throughput and continuous applications. To address these concerns, we have developed two microfluidic lobe filters bioinspired by the filtration mechanism of two species of manta ray. These chips enable filtration of particles around 10-30 μm with precise control and high throughput by using two arrays of equally spaced filter lobes. For each filter design, we investigated multiple inlet flow rates and particle sizes to identify successful operational parameters. Filtration efficiency increases with fluid flow rate, suggesting that particle inertial effects play a key role in lobe filter separation. Microparticle filtration efficiencies up to 99% were obtainable with inlet flow rates of
SUBMITTER: Clark AS
PROVIDER: S-EPMC8486309 | biostudies-literature | 2021 Sep
REPOSITORIES: biostudies-literature