Assessment of direct analysis in real time accurate mass spectrometry for the determination of triclosan in complex matrices.


ABSTRACT: In this work, the applicability of direct analysis in real time coupled to accurate mass spectrometry (DART-MS) to the quantitative determination of triclosan (TCS) in samples with increasing complexity, from personal care products to extracts from sewage, is investigated. In the first term, DART-MS spectra of TCS as free phenol and as derivatized species are characterized; thereafter, the effects of several instrumental variables in the detectability of TCS (i.e., temperature, solvent, and compound holder) are discussed. Under final selected conditions, TCS was determined from its [M-H]- ions, without need of derivatization, attaining an instrumental limit of quantification of 5 ng mL-1, with a linear response range up to 1000 ng mL-1. Complex matrices, such as solid-phase extracts obtained from environmental water samples, moderately inhibited the ionization efficiency of TCS, with signal attenuation percentages in the range of 6 to 57%, depending on the sample type and on the concentration factor provided by the SPE procedure. The accuracy of results obtained by DART-MS was evaluated using liquid chromatography (LC) with MS detection; in both cases, a time-of-flight (TOF) MS instrument was employed for the selective determination of the [M-H]- ions of TCS (m/z values 286.9439 and 288.9410) using a mass window of 20 ppm. DART-MS did not only provide enough sensitivity to detect the presence of TCS in environmental samples (raw and treated wastewater as well as freeze-dried sludge), but also measured concentrations matched those determined by LC-ESI-TOF-MS, with only slightly higher standard deviations. During analysis of personal care products, containing much higher concentrations of TCS in a less complex matrix, both techniques were equivalent in terms of accuracy and precision. Graphical abstract.

SUBMITTER: Cobo-Golpe M 

PROVIDER: S-EPMC8487875 | biostudies-literature | 2021 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Assessment of direct analysis in real time accurate mass spectrometry for the determination of triclosan in complex matrices.

Cobo-Golpe M M   García-Martín J J   Ramil M M   Cela R R   Rodríguez I I  

Analytical and bioanalytical chemistry 20210810 25


In this work, the applicability of direct analysis in real time coupled to accurate mass spectrometry (DART-MS) to the quantitative determination of triclosan (TCS) in samples with increasing complexity, from personal care products to extracts from sewage, is investigated. In the first term, DART-MS spectra of TCS as free phenol and as derivatized species are characterized; thereafter, the effects of several instrumental variables in the detectability of TCS (i.e., temperature, solvent, and comp  ...[more]

Similar Datasets

| S-EPMC7421040 | biostudies-literature
| S-EPMC6225280 | biostudies-literature
| S-EPMC10547761 | biostudies-literature
| S-EPMC8159215 | biostudies-literature
| S-EPMC4911641 | biostudies-literature
| S-EPMC9720406 | biostudies-literature
| S-EPMC3435450 | biostudies-literature
| S-EPMC2438217 | biostudies-literature
| S-EPMC6126382 | biostudies-literature
| S-EPMC5121580 | biostudies-literature