Technical considerations when using the EEG export of the SEDLine Root device.
Ontology highlight
ABSTRACT: Electroencephalographic (EEG) patient monitoring during general anesthesia can help to assess the real-time neurophysiology of unconscious states. Some monitoring systems like the SEDLine Root allow export of the EEG to be used for retrospective analysis. We show that changes made to the SEDLine display during recording affected the recorded EEG. These changes can strongly impact retrospective analysis of EEG signals. Real-time changes of the feed speed in the SEDLine Root device display modifies the sampling rate of the exported EEG. We used a patient as well as a simulated EEG recording to highlight the effects of the display settings on the extracted EEG. Therefore, we changed EEG feed and amplitude resolution on the display in a systematic manner. To visualize the effects of these chan
SUBMITTER: von Dincklage F