Method to measure the size of a radiographic field larger than a detector by imaging fluorescence X-rays with a slit camera.


ABSTRACT:

Purpose

X-ray imaging devices contain a collimator system that defines a rectangular irradiation field on the detector plane. The size and position of the X-ray field, and its congruence with the corresponding light field, must be regularly tested for quality control. We propose a new method to estimate how far the x-ray field extends beyond the detector which does not require the use of external detectors.

Methods

A metallic foil is inserted perpendicularly between the source and the detector. A slit camera, a linear extension of a pinhole camera, is used to project onto the detector the fluorescence X-rays emitted by the irradiated foil. The location where the fluorescence signal inside the camera vanishes is used to extrapolate the location of the field boundary. Monte Ca