Strain by virtual extensometers and video-imaging optical coherence tomography as a repeatable metric for IOP-Induced optic nerve head deformations.


ABSTRACT:

Purpose

To determine if in vivo strain response of the Optic Nerve Head (ONH) to IOP elevation visualized using Optical Coherence Tomography (OCT) video imaging and quantified using novel virtual extensometers was able to be provided repeatable measurements of tissue specific deformations.

Methods

The ONHs of 5 eyes from 5 non-human primates (NHPs) were imaged by Spectralis OCT. A vertical and a horizontal B-scan of the ONH were continuously recorded for 60 s at 6 Hz (video imaging mode) during IOP elevation from 10 to 30 mmHg. Imaging was repeated over three imaging sessions. The 2D normal strain was computed by template-matching digital image correlation using virtual extensometers. ANOVA F-test (F) was used to compare inter-eye, inter-session, and inter-tissue variability