Comparison and optimization of various moving patient-based real-time quality control procedures for serum sodium.


ABSTRACT:

Background

Patient-based real-time quality control (PBRTQC) is a valuable tool for monitoring the performance of testing processes. We aimed to compare and optimize various PBRTQC procedures for serum sodium.

Methods

In a computer simulation, artificial errors were added to 680,000 real patients' results. The characteristics of error detection of various algorithms-moving average, moving median, moving SD and moving proportion of normal results including different control limits (CLs)-were assessed on their ability to detect critical errors early.

Results

The moving average and moving median were sensitive to system error, and the moving SD tended to detect random error. P3SD (moving proportion of normal results, CLs based on mean and SD of proportion of no