Unknown

Dataset Information

Clinical evaluation of the automated Abbott RealTime SARS-CoV-2, Alinity m SARS-CoV-2, and Alinity m Resp-4-Plex assays.


ABSTRACT:

Background

Detection of SARS-CoV-2 infections relies on the use of sensitive, accurate and high throughput RT-PCR assays.

Objectives

We assessed the analytical performance of the Abbott RealTime SARS-CoV-2 (RT-SARS), Alinity m SARS-CoV-2 (AlinSARS) assays and compared the clinical performance of the RT-SARS, AlinSARS, and Alinity m Resp-4-Plex (Alin4Plex) assays to the Seegene Allplex assay (Allplex) and an inhouse test (Inhouse).

Results

We found 100 % positive percent agreement (PPA) and 100 % negative percent agreement (NPA) comparing RT-SARS and Allplex. RT-SARS, AlinSARS and Inhouse showed 100 % NPA and 100 % PPA across all assays, except for the RdRp target of Inhouse (PPA = 84 %). Similarly, Alin4Plex and Allplex showed high agreement with specimens containing

SUBMITTER: Ehret R 

PROVIDER: S-EPMC8532389 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature