An Automated Hydrodynamically Mediated Technique for Preparation of Calibration Solutions via Capillary Electrophoresis System as a Promising Alternative to Manual Pipetting.


ABSTRACT: In this paper, a novel procedure for preparing calibration solutions for capillary electrophoresis (CE)-based quantitative analysis is proposed. Our approach, named the automated hydrodynamically mediated technique (AHMT), uses a capillary and a pressure system to deliver the expected amount of working solution and diluent directly to a sample vial. As a result, calibration solutions are prepared automatically inside the CE instrument, without any or with minimal manual operation. Two different modes were tested: forward and reverse, differing in the direction of hydrodynamic flow. The calibration curves obtained for a model mixture of analytes using AHMT were thorough compared to the standard procedure based on manual pipetting. The results were consistent, though the volume of obtained calibration solutions and the potential risk of random errors were significantly minimized by AHMT. Its effectiveness was further enhanced by the application of SCIEX® nanoVials, reducing the actual volume of calibration solutions down to 10 μL.

SUBMITTER: Golab M 

PROVIDER: S-EPMC8540481 | biostudies-literature | 2021 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

An Automated Hydrodynamically Mediated Technique for Preparation of Calibration Solutions via Capillary Electrophoresis System as a Promising Alternative to Manual Pipetting.

Gołąb Małgorzata M   Woźniakiewicz Michał M   Nowak Paweł M PM   Kościelniak Paweł P  

Molecules (Basel, Switzerland) 20211016 20


In this paper, a novel procedure for preparing calibration solutions for capillary electrophoresis (CE)-based quantitative analysis is proposed. Our approach, named the automated hydrodynamically mediated technique (AHMT), uses a capillary and a pressure system to deliver the expected amount of working solution and diluent directly to a sample vial. As a result, calibration solutions are prepared automatically inside the CE instrument, without any or with minimal manual operation. Two different  ...[more]

Similar Datasets

| S-EPMC10569977 | biostudies-literature
| S-EPMC7852844 | biostudies-literature
| S-EPMC3622270 | biostudies-literature
| S-EPMC4097815 | biostudies-literature
| S-EPMC3458716 | biostudies-literature
| S-EPMC3257135 | biostudies-literature
| S-EPMC310688 | biostudies-literature
| S-EPMC6886162 | biostudies-literature