Periodic Comparability Verification and Within-Laboratory Harmonization of Clinical Chemistry Laboratory Results at a Large Healthcare Center With Multiple Instruments.


ABSTRACT:

Background

Results from laboratories using multiple instruments should be standardized or harmonized and comparability-verified for consistent quality control. We developed a simple frequent comparability verification methodology applicable to large healthcare centers using multiple clinical chemistry instruments from different manufacturers.

Methods

Comparability of five clinical chemistry instruments (Beckman Coulter AU5800, Abbott Architect Ci16000, two Siemens Vista 1500, and Ortho Vitros 5600) was evaluated from 2015 to 2019 for 12 clinical chemistry measurements. Pooled residual patient samples were used for weekly verifications. Results from any instrument exceeding the allowable verification range versus the results from the comparative instrument (AU5800) were repor