Dosimetry of a novel converging X-ray source for kilovoltage radiotherapy.


ABSTRACT:

Purpose

The objective of this work was to evaluate phantom dosimetry of a novel kilovoltage (kV) X-ray source, which employs a stationary tungsten anode and a linearly swept scanning electron beam. The source utilizes converging X-ray collimation along with orthogonal mechanical rotation to distribute surface flux over large area. In this study, this was investigated as a potential solution to fast-falloff limitations expected with kV radiotherapy. This was done with the aim of future clinical development of a lower cost radiotherapy alternative to megavoltage (MV) linac systems.

Methods

Radiochromic film was employed for dosimetry on the kV X-ray source of the linear-converging radiotherapy system (LCRS). The source utilizes charge particle optics to magnetically deflect an

SUBMITTER: Stalbaum T 

PROVIDER: S-EPMC8568629 | biostudies-literature | 2021 Oct

REPOSITORIES: biostudies-literature