Manual QT interval measurement with a smartphone-operated single-lead ECG versus 12-lead ECG: a within-patient diagnostic validation study in primary care.


ABSTRACT:

Objective

To determine the accuracy of QT measurement in a smartphone-operated, single-lead ECG (1L-ECG) device (AliveCor KardiaMobile 1L).

Design

Cross-sectional, within-patient diagnostic validation study.

Setting/participants

Patients underwent a 12-lead ECG (12L-ECG) for any non-acute indication in primary care, April 2017-July 2018.

Intervention

Simultaneous recording of 1L-ECGs and 12L-ECGs with blinded manual QT assessment. OUTCOMES OF INTEREST: (1) Difference in QT interval in milliseconds (ms) between the devices; (2) measurement agreement between the devices (excellent agreement <20 ms and clinically acceptable agreement <40 ms absolute difference); (3) sensitivity and specificity for detection of extreme QTc (short (≤340 ms) or long (≥480 ms)), on 1L