Unknown

Dataset Information

Evaluation of Robustness of Local Phase Velocity Imaging in Homogenous Tissue-Mimicking Phantoms.


ABSTRACT: Shear wave elastography (SWE) is a method of evaluating mechanical properties of soft tissues. Most current implementations of SWE report the group velocity for shear wave velocity, which assumes an elastic, isotropic, homogenous and incompressible tissue. Local phase velocity imaging (LPVI) is a novel method of phase velocity reconstruction that allows for accurate evaluation of shear wave velocity at specified frequencies. This method's robustness was evaluated in 11 elastic and 8 viscoelastic phantoms using linear and curvilinear arrays. We acquired data with acoustic radiation force push beams with different focal depths and F-numbers and reconstructed phase velocity images over a wide range of frequencies. Regardless of phantom, push beam focal depth and reconstruction frequency, an F

SUBMITTER: Wood BG 

PROVIDER: S-EPMC8578323 | biostudies-literature | 2021 Dec