Ultrasound Cyclo Plasty for Treatment of Surgery-Naive Open-Angle Glaucoma Patients: A Prospective, Multicenter, 2-Year Follow-Up Trial.


ABSTRACT: The purpose of this prospective study was to evaluate the efficacy and safety of the Ultrasound Cyclo Plasty (UCP) procedure using high-intensity focused ultrasound in surgery-naïve open-angle glaucoma patients. prospective, non-randomized, single-arm, multicenter clinical trial. Sixty-six eyes with primary open-angle glaucoma, intraocular pressure (IOP) ≥21 mmHg and with no history of filtering surgery were enrolled. Patients were treated by UCP with a therapy probe comprising six piezoelectric transducers, consecutively activated for 8 s each. Complete ophthalmic examination was performed before the procedure, 1 day after the procedure, and 1, 3, 6, 12, 18 and 24 months after the procedure. Primary outcomes were complete success (defined as IOP lowering from baseline ≥20% without additio