Automated Assessment of the Quality of Phonocardographic Recordings through Signal-to-Noise Ratio for Home Monitoring Applications.
Ontology highlight
ABSTRACT: The signal quality limits the applicability of phonocardiography at the patients' domicile. This work proposes the signal-to-noise ratio of the recorded signal as its main quality metrics. Moreover, we define the minimum acceptable values of the signal-to-noise ratio that warrantee an accuracy of the derived parameters acceptable in clinics. We considered 25 original heart sounds recordings, which we corrupted by adding noise to decrease their signal-to-noise ratio. We found that a signal-to-noise ratio equal to or higher than 14 dB warrants an uncertainty of the estimate of the valve closure latencies below 1 ms. This accuracy is higher than that required by most clinical applications. We validated the proposed method against a public database, obtaining results comparable to those obtain