Catheter Port Reversal in Citrate Continuous Veno-Venous Hemofiltration.


ABSTRACT:

Introduction

Knowledge of effects of catheter port reversal (CathPR), when blood is withdrawn from the venous port and returned via the arterial port, often used in dysfunctional catheters in renal replacement therapy, is limited in the setting of citrate continuous veno-venous hemofiltration (CVVH).

Methods

In this open trial, post-filter ionized calcium (PfiCa), post-filter citrate concentration (PfCC), catheter recirculation, and solute clearance were measured before, during, and after 6 hours of CathPR, in well-functioning catheters. All other settings, including citrate settings, were left constant during the study.

Results

Twenty-three patients were included. Mean PfiCa before CathPR of 0.36 mmol/L (SD 0.06) decreased to 0.31 (0.04) after 2 hours (P = 0.0

SUBMITTER: Boer W 

PROVIDER: S-EPMC8589722 | biostudies-literature | 2021 Nov