Enumeration of viable CD34+ cells in cord blood using a novel stem cell enumeration kit.


ABSTRACT:

Objective

To assess the detection performance of the hematopoietic stem cell enumeration kit developed by BD Biosciences.

Methods

Cord blood samples were prepared using a hematopoietic stem cell enumeration kit developed by BD Biosciences and Stem-Kit reagents from Beckman Coulter. CD34+ cells were enumerated using a BD FACSCanto instrument and FACSDiva software.

Results

A total of 519 samples were analyzed in this study. The hematopoietic stem cell enumeration kit developed by BD Biosciences yielded absolute counts of CD34-positive cells that were on average 8.7% lower than Beckman Coulter Stem-Kit reagents (range: -5.7% to-14.7%). The BD Biosciences kit yielded relative counts that were on average 9.9% higher compared with Beckman Coulter Stem-Kit reagents (range: -2.1% to +13.8%). The intraclass correlation coefficients for absolute and relative counts of CD34-positive cells were 0.9967 (95% confidence interval [CI]: 0.9961-0.9972) and 0.9512 (95% CI: 0.9423-0.9587) for the BD Biosciences and Beckman Coulter kits, respectively.

Conclusions

The hematopoietic stem cell enumeration kit developed by BD Biosciences can be used to enumerate CD34-positive stem cells from cord blood samples.

SUBMITTER: Huang S 

PROVIDER: S-EPMC8593314 | biostudies-literature | 2021 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Enumeration of viable CD34+ cells in cord blood using a novel stem cell enumeration kit.

Huang Sheng S   Xu Kangli K   Xu Shanshan S   Yu Shijun S   Miao Yan Y   Wang Juan J  

The Journal of international medical research 20211101 11


<h4>Objective</h4>To assess the detection performance of the hematopoietic stem cell enumeration kit developed by BD Biosciences.<h4>Methods</h4>Cord blood samples were prepared using a hematopoietic stem cell enumeration kit developed by BD Biosciences and Stem-Kit reagents from Beckman Coulter. CD34+ cells were enumerated using a BD FACSCanto instrument and FACSDiva software.<h4>Results</h4>A total of 519 samples were analyzed in this study. The hematopoietic stem cell enumeration kit develope  ...[more]

Similar Datasets

| S-EPMC6275752 | biostudies-literature
| S-EPMC8885393 | biostudies-literature
| S-EPMC6400716 | biostudies-literature
| S-EPMC4479625 | biostudies-literature
| S-EPMC4161396 | biostudies-literature
| S-EPMC4799819 | biostudies-literature
| S-EPMC5989201 | biostudies-literature
| S-EPMC10845921 | biostudies-literature
| S-EPMC2703252 | biostudies-literature
| S-EPMC9929769 | biostudies-literature