Determining the longitudinal accuracy and reproducibility of T1 and T2 in a 3T MRI scanner.


ABSTRACT:

Purpose

To determine baseline accuracy and reproducibility of T1 and T2 relaxation times over 12 months on a dedicated radiotherapy MRI scanner.

Methods

An International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) System Phantom was scanned monthly on a 3T MRI scanner for 1 year. T1 was measured using inversion recovery (T1 -IR) and variable flip angle (T1 -VFA) sequences and T2 was measured using a multi-echo spin echo (T2 -SE) sequence. For each vial in the phantom, accuracy errors (%bias) were determined by the relative differences in measured T1 and T2 times compared to reference values. Reproducibility was measured