Ontology highlight
ABSTRACT: Purpose
To determine baseline accuracy and reproducibility of T1 and T2 relaxation times over 12 months on a dedicated radiotherapy MRI scanner.Methods
An International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) System Phantom was scanned monthly on a 3T MRI scanner for 1 year. T1 was measured using inversion recovery (T1 -IR) and variable flip angle (T1 -VFA) sequences and T2 was measured using a multi-echo spin echo (T2 -SE) sequence. For each vial in the phantom, accuracy errors (%bias) were determined by the relative differences in measured T1 and T2 times compared to reference values. Reproducibility was measured
SUBMITTER: Carr ME
PROVIDER: S-EPMC8598150 | biostudies-literature | 2021 Nov
REPOSITORIES: biostudies-literature