Analytical and clinical evaluation of a novel assay for measurement of interleukin 6 in human whole blood samples.


ABSTRACT:

Background

Interleukin 6 assays are useful in early detection of infections and risk stratification of critically ill patients, so an assay with a short turnaround-time and near-patient use is preferred. This study evaluated the performance of a new interleukin 6 assay, Pylon IL-6 assay, and explored its potential use in near-patient settings.

Methods

We carried out imprecision, linearity and comparison studies using serum and plasma samples according to CLSI EP guidelines. The stability of whole blood samples during storage was assessed. Furthermore, whole blood samples from pediatric patients with suspected infection were measured to evaluate the assay's diagnostic performance.

Results

The within-run CVs and total CVs of Pylon IL-6 assay were determined as 1.8% and