Unknown

Dataset Information

Magnetic field compensation coil design for magnetoencephalography.


ABSTRACT: While optically pumped magnetometers (OPMs) can be attached to the head of a person and allow for highly sensitive recordings of the human magnetoencephalogram (MEG), they are mostly limited to an operational range of approximately 5 nT. Consequently, even inside a magnetically shielded room (MSR), movements in the remnant magnetic field disable the OPMs. Active suppression of the remnant field utilizing compensation coils is therefore essential. We propose 8 compensation coils on 5 sides of a cube with a side length of approximately 2 m which were optimized for operation inside an MSR. Compared to previously built bi-planar compensation coils, the coils proposed in this report are more complex in geometry and achieved smaller errors for simulated compensation fields. The proposed coils wi

SUBMITTER: Kutschka H 

PROVIDER: S-EPMC8608906 | biostudies-literature | 2021 Nov

REPOSITORIES: biostudies-literature