Ontology highlight
ABSTRACT: Objective
The aim of this study was to evaluate the effect of field of view (FOV) and beam energy on the scatter-to-primary ratio (SPR) in dental cone-beam CT (CBCT).Methods
An anthropomorphic phantom representing an adult male (ATOM Max 711-HN, Norfolk, VA, USA) was scanned using the 3D Accuitomo 170 CBCT (J. Morita, Kyoto, Japan) using 11 FOVs. During each scan, half of the X-ray beam was blocked. Each scan was performed at three exposure settings with varying beam energy and equal radiation dose: 90 kV 5 mA, 77 kV 7.5 mA and 69 kV 10 mA. The SPR was estimated by measuring the grey values in the blocked and non-blocked regions of the RAW data. The effect of FOV on SPR was evaluated using Dunn's multiple comparison test, and the effect of the exposure settings was compared
SUBMITTER: Pauwels R
PROVIDER: S-EPMC8611284 | biostudies-literature | 2021 Dec
REPOSITORIES: biostudies-literature