Microparticle-tagged image-based cell counting (ImmunoSpin) for CD4 + T cells.


ABSTRACT: Affordable point-of-care (POC) CD4 + T lymphocyte counting techniques have been developed as alternatives to flow cytometry-based instruments caring for patients with human immunodeficiency virus (HIV)-1. However, POC CD4 enumeration technologies can be inaccurate. Here, we developed a microparticle-based visual detector of CD4 + T lymphocytes (ImmunoSpin) using microparticles conjugated with anti-CD4 antibodies, independent of microfluidic or fluorescence detection systems. Visual enumeration of CD4 + T cells under conventional light microscope was accurate compared to flow cytometry. Microparticle-tagged CD4 + T cells were well-recognized under a light microscope. ImmunoSpin showed very good precision (coefficients of variation of ImmunoSpin were ≤ 10%) and high correlation with clinical-grade flow cytometry for the enumeration of CD4 + T cells (y = 0.4232 + 0.9485 × for the %CD4 + T cell count, R2 = 0.99). At thresholds of 200 and 350 cells/µL, there was no misclassification of the ImmunoSpin system compared to the reference flow cytometry. ImmunoSpin showed clear differential classification of CD4 + T lymphocytes from granulocytes and monocytes. Because non-fluorescence microparticle-tags and cytospin slides are used in ImmunoSpin, they can be applied to an automatic digital image analyzer. Slide preparation allows long-term storage, no analysis time limitations, and image transfer in remote areas.

SUBMITTER: Hwang SH 

PROVIDER: S-EPMC8616869 | biostudies-literature | 2021 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Microparticle-tagged image-based cell counting (ImmunoSpin) for CD4 + T cells.

Hwang Sang-Hyun SH   Yang John Jeongseok JJ   Oh Yoon-Hee YH   Ko Dae-Hyun DH   Sung Heungsup H   Cho Young-Uk YU   Jang Seongsoo S   Park Chan-Jeoung CJ   Oh Heung-Bum HB  

Mikrochimica acta 20211125 12


Affordable point-of-care (POC) CD4 + T lymphocyte counting techniques have been developed as alternatives to flow cytometry-based instruments caring for patients with human immunodeficiency virus (HIV)-1. However, POC CD4 enumeration technologies can be inaccurate. Here, we developed a microparticle-based visual detector of CD4 + T lymphocytes (ImmunoSpin) using microparticles conjugated with anti-CD4 antibodies, independent of microfluidic or fluorescence detection systems. Visual enumeration o  ...[more]

Similar Datasets

| S-EPMC10871079 | biostudies-literature
| S-EPMC3524557 | biostudies-literature
| S-EPMC8719728 | biostudies-literature
| S-EPMC9849527 | biostudies-literature
| S-EPMC7249160 | biostudies-literature
| S-EPMC4746068 | biostudies-literature
| S-EPMC3316693 | biostudies-other
| S-EPMC4537305 | biostudies-literature
| S-EPMC10096914 | biostudies-literature
| S-EPMC8374203 | biostudies-literature