ABSTRACT: Purpose
Long-leg-radiography (LLR) is commonly used for the measurement of lower limb alignment. However, limb rotations during radiography may interfere with the alignment measurement. This study examines the effect of limb rotation on the accuracy of measurements based on the mechanical and anatomical axes of the femur and tibia, with variations in knee flexion and coronal deformity.Methods
Forty-five lower limbs of 30 patients were scanned with CT. Virtual LLRs simulating five rotational positions (neutral, ± 10[Formula: see text], and ± 20[Formula: see text] internal rotation) were generated from the CT images. Changes in the hip-knee-ankle angle (HKA) and the femorotibial angle (FTA) were measured on each image with respect to neutral values. These changes were related