ABSTRACT: Objectives
This study aims to evaluate the accuracy of physical density prediction in single-energy CT (SECT) and dual-energy CT (DECT) by adapting a fully simulation-based method using a material-based forward projection algorithm (MBFPA).Methods
We used biological tissues referenced in ICRU Report 44 and tissue substitutes to prepare three different types of phantoms for calibrating the Hounsfield unit (HU)-to-density curves. Sinograms were first virtually generated by the MBFPA with four representative energy spectra (i.e. 80 kVp, 100 kVp, 120 kVp, and 6 MVp) and then reconstructed to form realistic CT images by adding statistical noise. The HU-to-density curves in each spectrum and their pairwise combinations were derived from the CT images. The accuracy of these