Signal-to-noise ratio (SNR) is used to evaluate the performance of magnetic resonance (MR) imaging systems. Accurate and consistent estimations are needed to routinely use SNR to assess coils and image reconstruction techniques.Purpose
To identify a reliable and practical method for SNR estimation in multiple-coil reconstructions.Study type
Technical evaluation and comparison.Subjects/phantom
A uniform phantom and four healthy volunteers: 35, 38, 39 y/o males, 25 y/o female.Field strength/sequence
Two-dimensional multislice gradient-echo pulse sequence at 3 T and 7 T.Assessment
Reference-standard SNR was calculated from 100 multiple replicas. Six SNR methods were compared against it: difference image (DI), analytic array combinatio