Potentiometric pH Nanosensor for Intracellular Measurements: Real-Time and Continuous Assessment of Local Gradients.
Ontology highlight
ABSTRACT: We present a pH nanosensor conceived for single intracellular measurements. The sensing architecture consisted of a two-electrode system evaluated in the potentiometric mode. We used solid-contact carbon nanopipette electrodes tailored to produce both the indicator (pH nanosensor) and reference electrodes. The indicator electrode was a membrane-based ion-selective electrode containing a receptor for hydrogen ions that provided a favorable selectivity for intracellular measurements. The analytical features of the pH nanosensor revealed a Nernstian response (slope of -59.5 mV/pH unit) with appropriate repeatability and reproducibility (variation coefficients of <2% for the calibration parameters), a fast response time (<5 s), adequate medium-term drift (0.7 mV h-1), and a linear r