Unknown

Dataset Information

The ACURATE neo2 valve system for transcatheter aortic valve implantation: 30-day and 1-year outcomes.


ABSTRACT:

Background

Transcatheter aortic valve implantation (TAVI) has become standard treatment for elderly patients with symptomatic severe aortic valve stenosis. The ACURATE neo AS study evaluates 30-day and 1-year clinical and hemodynamic outcomes in patients treated with the ACURATE neo2 valve.

Methods

The primary endpoint of this single-arm multicenter study is 30-day all-cause mortality. Other key endpoints include device performance, echocardiographic measures assessed by an independent core laboratory, and VARC-2 clinical efficacy and safety endpoints through 12 months.

Results

The study enrolled 120 patients (mean age 82.1 ± 4.0 years; 67.5% female, mean baseline STS score 4.8 ± 3.8%). The VARC-2 composite safety endpoint at 30 days occurred in 13.3% of patients. All

SUBMITTER: Mollmann H 

PROVIDER: S-EPMC8639565 | biostudies-literature | 2021 Dec

REPOSITORIES: biostudies-literature