Development and Provisional Validation of a Multiplex LC-MRM-MS Test for Timely Kidney Injury Detection in Urine.
Ontology highlight
ABSTRACT: Kidney injury is a complication frequently encountered in hospitalized patients. Early detection of kidney injury prior to loss of renal function is an unmet clinical need that should be targeted by a protein-based biomarker panel. In this study, we aim to quantitate urinary kidney injury biomarkers at the picomolar to nanomolar level by liquid chromatography coupled to tandem mass spectrometry in multiple reaction monitoring mode (LC-MRM-MS). Proteins were immunocaptured from urinary samples, denatured, reduced, alkylated, and digested into peptides before LC-MRM-MS analysis. Stable-isotope-labeled peptides functioned as internal standards, and biomarker concentrations were attained by an external calibration strategy. The method was evaluated for selectivity, carryover, matrix effects, l
SUBMITTER: van Duijl TT
PROVIDER: S-EPMC8650098 | biostudies-literature | 2021 Dec
REPOSITORIES: biostudies-literature
ACCESS DATA