A compact and mobile hybrid C-arm scanner for simultaneous nuclear and fluoroscopic image guidance.


ABSTRACT:

Purpose

This study evaluates the performance of a mobile and compact hybrid C-arm scanner (referred to as IXSI) that is capable of simultaneous acquisition of 2D fluoroscopic and nuclear projections and 3D image reconstruction in the intervention room.

Results

The impact of slightly misaligning the IXSI modalities (in an off-focus geometry) was investigated for the reduction of the fluoroscopic and nuclear interference. The 2D and 3D nuclear image quality of IXSI was compared with a clinical SPECT/CT scanner by determining the spatial resolution and sensitivity of point sources and by performing a quantitative analysis of the reconstructed NEMA image quality phantom. The 2D and 3D fluoroscopic image of IXSI was compared with a clinical CBCT scanner by visualizing the Fluorad A+D image quality phantom and by visualizing a reconstructed liver nodule phantom. Finally, the feasibility of dynamic simultaneous nuclear and fluoroscopic imaging was demonstrated by injecting an anthropomorphic phantom with a mixture of iodinated contrast and 99mTc.

Conclusion

Due to the divergent innovative hybrid design of IXSI, concessions were made to the nuclear and fluoroscopic image qualities. Nevertheless, IXSI realizes unique image guidance that may be beneficial for several types of procedures.

Key points

• IXSI can perform time-resolved planar (2D) simultaneous fluoroscopic and nuclear imaging. • IXSI can perform SPECT/CBCT imaging (3D) inside the intervention room.

SUBMITTER: Dietze MMA 

PROVIDER: S-EPMC8660732 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

A compact and mobile hybrid C-arm scanner for simultaneous nuclear and fluoroscopic image guidance.

Dietze Martijn M A MMA   Kunnen Britt B   Brontsema Frank F   Ramaekers Pascal P   Beijst Casper C   Afifah Maryam M   Braat Arthur J A T AJAT   Lam Marnix G E H MGEH   de Jong Hugo W A M HWAM  

European radiology 20210616 1


<h4>Purpose</h4>This study evaluates the performance of a mobile and compact hybrid C-arm scanner (referred to as IXSI) that is capable of simultaneous acquisition of 2D fluoroscopic and nuclear projections and 3D image reconstruction in the intervention room.<h4>Results</h4>The impact of slightly misaligning the IXSI modalities (in an off-focus geometry) was investigated for the reduction of the fluoroscopic and nuclear interference. The 2D and 3D nuclear image quality of IXSI was compared with  ...[more]