Ontology highlight
ABSTRACT: Purpose
To evaluate a two-dimensional diode array for patient-specific quality assurance of VMAT stereotactic radiosurgery (SRS) plans.Methods
The diode array, an SRS MapCHECK (SRSMC), was composed of a 77 mm ×77 mm face-centered array having a spacing of 2.47 mm. Sixty SRS plans were selected from our clinical database, 30 for treatment of a single target and 30 for multiple targets. The target sizes ranged from 2.4 mm to 44.7 mm equivalent diameter (median 8.7 mm). The plans were delivered to the diode array. For multiple target plans, two measurements were obtained at two locations, one corresponding to the largest target and the other to the smallest target. Gamma using a 3%/1 mm criteria and the dose to the center diode were compared with radiochromic film (RCF). Dose
SUBMITTER: Popple RA
PROVIDER: S-EPMC8664143 | biostudies-literature | 2021 Dec
REPOSITORIES: biostudies-literature